Abiomed, Inc. Recall
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA;
Updated February 19, 2019
Abiomed, Inc. is recalling Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA;. Hazard: Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
Is the Abiomed, Inc. recalled? Yes. This product was recalled by FDA on February 19, 2019. Details below.
What is recalled
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
Am I affected
You may be affected if you own this product. Check the brand, model, and lot/date codes listed in the recall against your item.
What to do right now
Voluntary: Firm initiated
The hazard
Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
Brands affected
- Source
- FDA
- Category
- Drugs & Medications
- Recall ID
- fda-Z-3251-2026
- Classification
- Class II
- Recalled
- February 19, 2019
Get free recall alerts
New Drugs & Medications recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Drugs & Medications recalls

ScotchTM Thermal Laminators Recall
ScotchTM Thermal Laminators is recalling 3M Company Recalls ScotchTM Thermal Laminators Due to Risk of Serious Injury or Death from Fire Hazard. Hazard: The …
Scotch™ Thermal Laminators

The manufacturer Recall
The manufacturer is recalling Regfear Recalls BedsPick Mattresses Due to Risk of Serious Injury or Death from Fire; Violate Mandatory Flammability Standard for …
BedsPick Mattresses

BedsPick Mattresses Recall
BedsPick Mattresses is recalling Regfear Recalls BedsPick Mattresses Due to Risk of Serious Injury or Death from Fire; Violate Mandatory Flammability Standard …
BedsPick Mattresses