High risk Baby & KidsClass II

GE CARESCAPE Telemetry Server Recall

GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih Ap

Updated April 24, 2026

The GE HealthCare CARESCAPE Telemetry Server ARK-2250L may stop monitoring patient heart rhythm and oxygen levels if the system loses network connection. This loss of critical monitoring data could delay detection of dangerous heart conditions.

Is the GE Medical Systems Information Technologies Inc recalled? Yes. This product was recalled by FDA on April 24, 2026. Details below.

What is recalled

GE HealthCare CARESCAPE Telemetry Server ARK-2250L (NORTH AMERICA), REF #5697603-902, Product Code KSA, used with ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0.

Am I affected

You're affected if your hospital or care facility uses the GE HealthCare CARESCAPE Telemetry Server ARK-2250L (reference #5697603-902, product code KSA) connected to ApexPro CTS v5.0 or v6.0.

What to do right now

Contact your GE HealthCare representative or facility biomedical team immediately to assess whether your system is affected and obtain any necessary software updates or operational guidance to prevent loss of patient monitoring.

The hazard

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Brands affected

Source
FDA
Category
Baby & Kids
Recall ID
fda-Z-2246-2026
Classification
Class II
Recalled
April 24, 2026
Read the official notice

Get free recall alerts

New Baby & Kids recalls, the moment they post.

Free forever. Unsubscribe anytime. We never sell your data.