Medium risk Drugs & MedicationsClass II

Intuitive Surgical, Inc. Recall

Da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Updated December 30, 2025

Intuitive Surgical, Inc. is recalling da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45. Hazard: Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Is the Intuitive Surgical, Inc. recalled? Yes. This product was recalled by FDA on December 30, 2025. Details below.

What is recalled

Da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Am I affected

You may be affected if you own this product. Check the brand, model, and lot/date codes listed in the recall against your item.

What to do right now

Voluntary: Firm initiated

The hazard

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Brands affected

Source
FDA
Category
Drugs & Medications
Recall ID
fda-Z-1355-2026
Classification
Class II
Recalled
December 30, 2025
Read the official notice

Get free recall alerts

New Drugs & Medications recalls, the moment they post.

Free forever. Unsubscribe anytime. We never sell your data.