Medium risk Drugs & MedicationsClass II

Magellan Diagnostics, Inc. Recall

LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 c

Updated March 13, 2026

Magellan Diagnostics, Inc. is recalling LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 c. Hazard: Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Is the Magellan Diagnostics, Inc. recalled? Yes. This product was recalled by FDA on March 13, 2026. Details below.

What is recalled

LeadCare¿ Plus Blood Lead Test Kit; Catalog Number: 82-0004; The LeadCare¿ Plus kit is provided with the following materials: " Sensors (4 containers of 24 each; 96 tests) " Treatment Reagent Tubes (4 packages of 24 each; 96 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)

Am I affected

You may be affected if you own this product. Check the brand, model, and lot/date codes listed in the recall against your item.

What to do right now

Voluntary: Firm initiated

The hazard

Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed results and additional unnecessary testing.

Brands affected

Source
FDA
Category
Drugs & Medications
Recall ID
fda-Z-1894-2026
Classification
Class II
Recalled
March 13, 2026
Read the official notice

Get free recall alerts

New Drugs & Medications recalls, the moment they post.

Free forever. Unsubscribe anytime. We never sell your data.