Medline Strykeflow 2 Suction Irrigator Recall
Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G
Updated December 16, 2025
Medline robotics urology prostate kits containing Stryker Strykeflow 2 Suction Irrigators may leak or emit vaporized saline during use due to a 2024 design change that allows irrigation solution to reach the handpiece and battery pack.
Is the Medline Industries, LP recalled? Yes. This product was recalled by FDA on December 16, 2025. Details below.
What is recalled
Medline medical procedure convenience kits labeled as KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G, containing Stryker Strykeflow 2 Suction Irrigators.
Am I affected
You're affected if you have a Medline robotics urology prostate kit (REF DYKMBNDL116G) that contains a Stryker Strykeflow 2 Suction Irrigator purchased or in use in 2024 or later.
What to do right now
Stop using the Strykeflow 2 Suction Irrigator in this kit immediately. Contact Medline Industries for instructions on how to proceed with the recall.
The hazard
Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-1090-2026
- Classification
- Class II
- Recalled
- December 16, 2025
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls
ChloraPrep FREPP Applicators Recall
ChloraPrep FREPP sterile skin antiseptic applicators were distributed without proper sterilization, posing a risk of infection. These applicators are used …
CareFusion 213, LLC
Aquacel Ag Surgical SP Recall
ConvaTec is recalling Aquacel Ag Surgical SP surgical dressings (REF 413556, 9cm x 15cm size) because some were made thicker than the design specification. The …
ConvaTec, Inc

SUNS International SS6291 Elevator Switch Recall
SUNS International is recalling SS6291 Solenoid Interlock Switches used in residential elevators because the switch can get stuck in a retracted position, …
SUNS SS6291 Solenoid Interlock Switches