Philips North America Recall
Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions:
Updated August 28, 2026
Philips North America is recalling Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions:. Hazard: When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Is the Philips North America recalled? Yes. This product was recalled by FDA on August 28, 2026. Details below.
What is recalled
Ingenia 3.0T and 3.0T CX; Model number (REF): 1. 781342 (3.0T), 2. 781377 (3.0T), 3. 782103 (3.0T), 4. 781271 (3.0T CX); Software Versions: R11, R12 and R13;
Am I affected
You may be affected if you own this product. Check the brand, model, and lot/date codes listed in the recall against your item.
What to do right now
Voluntary: Firm initiated
The hazard
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Brands affected
- Source
- FDA
- Category
- Drugs & Medications
- Recall ID
- fda-Z-3309-2026
- Classification
- Class II
- Recalled
- August 28, 2026
Get free recall alerts
New Drugs & Medications recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Drugs & Medications recalls

ScotchTM Thermal Laminators Recall
ScotchTM Thermal Laminators is recalling 3M Company Recalls ScotchTM Thermal Laminators Due to Risk of Serious Injury or Death from Fire Hazard. Hazard: The …
Scotch™ Thermal Laminators

The manufacturer Recall
The manufacturer is recalling Regfear Recalls BedsPick Mattresses Due to Risk of Serious Injury or Death from Fire; Violate Mandatory Flammability Standard for …
BedsPick Mattresses

BedsPick Mattresses Recall
BedsPick Mattresses is recalling Regfear Recalls BedsPick Mattresses Due to Risk of Serious Injury or Death from Fire; Violate Mandatory Flammability Standard …
BedsPick Mattresses